Intraoperative radiation therapy
Intraoperative radiation therapy | |
---|---|
Intervention | |
ICD-9-CM | 92.4 |
Intraoperative radiation therapy, or IORT is the application of therapeutic levels of radiation to the tumor bed while the area is exposed during surgery. IORT is typically a component in the multidisciplinary treatment of locally advanced and recurrent cancer, in combination with external beam radiation, surgery and chemotherapy. As a growing trend in recent years, IORT can also be used in earlier stage cancers such as prostate and breast cancer.
Medical uses
IORT was found to be useful and feasible in the multidisciplinary management of many solid tumors but further studies are needed to determine the benefit more precisely.[1] Single-institution experiences have suggested a role of IORT e.g. in brain tumors and cerebral metastases, locally advanced and recurrent rectal cancer, skin cancer, retroperitoneal sarcoma, pancreatic cancer and selected gynaecologic and genitourinary malignancies. For local recurrences, irradiation with IORT is besides brachytherapy the only radiotherapeutic option if repeated EBRT is no longer possible. Generally, the normal tissue tolerance does not allow a second full-dose course of EBRT, even after years.[2][3]
Breast cancer
On 25 July the UK National Institute for Health and Care Excellence (NICE) gave provisional recommendation for the use of TARGIT IORT with Intrabeam in the UK National Health Service.[4] The 2015 update of guidelines of the Association of Gynecological Oncology (AGO), an autonomous community of the German Society of Gynecology and Obstetrics (DGGG) and the German Cancer Society includes TARGIT IORT during lumpectomy as a recommended option for women with a T1, Grade 1 or 2, ER positive breast cancer.[5]
Rationale
The rationale for IORT is to deliver a high dose of radiation precisely to the targeted area with minimal exposure of surrounding tissues which are displaced or shielded during the IORT. Conventional radiation techniques such as external beam radiotherapy (EBRT) following surgical removal of the tumor have several drawbacks: The tumor bed where the highest dose should be applied is frequently missed due to the complex localization of the wound cavity even when modern radiotherapy planning is used. Additionally, the usual delay between the surgical removal of the tumor and EBRT may allow a repopulation of the tumor cells. These potentially harmful effects can be avoided by delivering the radiation more precisely to the targeted tissues leading to immediate sterilization of residual tumor cells. Another aspect is that wound fluid has a stimulating effect on tumor cells. IORT was found to inhibit the stimulating effects of wound fluid.[6]
Methods
There are several methods that are used to deliver IORT. IORT can be delivered using electron beams (electron IORT), orthovoltage (250-300 kV) x-rays (X-ray IORT), high-dose-rate brachytherapy (HDR-IORT), or low energy (50 kV) x-rays (low-energy IORT).
Electron IORT
While IORT was first used in clinical practice in 1905,[7][8] the modern era of IORT began with the introduction of electron IORT in the mid-1960s by transporting patients from the OR after the tumor was removed to the radiation department to receive their electron IORT.[9][10] Electron IORT has the advantages of being able to carefully control the depth of radiation penetration while providing a very uniform dose to the tumor bed. Applied with energies in the range of 3 MeV to 12 MeV, electron IORT can treat to depths of up to 4 cm over areas as large as 300 cm2 (i.e. a 10 cm diameter circle) and takes only 1–3 minutes to deliver the prescribed radiation dose. A few hospitals built shielded operation rooms in which a conventional linear accelerator was installed to deliver the IORT radiation. This eliminated the complex logistics involved with patient transportation, but was so costly that only a few hospitals were able to use this approach. The breakthrough came in 1997, with the introduction of a miniaturized, self-shielded mobile-linear accelerator (Mobetron, IntraOp Corporation, US) [11] and a mobile but unshielded linear accelerator (Novac, Liac—SIT, Italy). More than 75,000 patients have been treated with electron IORT, almost half of them since the introduction of mobile electron IORT technology.
X-Ray IORT
Early practitioners of IORT treated primarily abdominal malignancies using superficial x-rays (75-125 kV) and later orthovoltage x-rays (up to 300 kV in energy) prior to the advent of technology that enabled high energy electrons. For the first 75 years, X-Ray IORT was used mostly for palliation, but there were a few anecdotal reports of long-term survivors. In the early 1980s, when the use of electron IORT was increasing and showed promising results for certain indications, a handful of hospitals installed othovoltage units in lightly shielded ORs to see if this lower cost approach could achieve comparable results to that of electron IORT. This approach was less costly than building a shielded OR for an electron IORT unit and eliminated the logistics involved with patient transportation. However, it had a number of problems that limited its appeal. X-Ray IORT has a poor uniformity of dose as a function of depth of penetration, the radiation does not stop at a pre-defined depth but continues to deposit radiation to underlying structures, and can do damage to boney structures if too high a dose is delivered. Despite its long use (since the 1930s), less than 1000 patients have been treated with this approach, and it is no longer offered at most centers.[12]
HDR-IORT
This technique was developed in the late-1980s in an attempt to combine the dosimetric advantages of high-dose rate brachytherapy with the challenges of treating some complex anatomic surfaces with IORT. It has the advantage of being lower cost than dedicated electron IORT systems, since many radiation centers already have an HDR system that can be transported to the OR when HDR-IORT is needed. HDR-IORT can also treat very large and convoluted surfaces. However, it does require a shielded OR or a shielded room in the OR complex to deliver the HDR-IORT.[13] The depth of penetration is very limited, typically either ½ cm to 1 cm depth, sometimes requiring extensive surgery due to the limited penetration of the radiation. Treatments tend to be 40 minutes or longer, resulting in greater OR time, more anesthesia and greater blood loss when compared to electron IORT. There are about 10 to 20 active centers using HDR-IORT for locally advanced and recurrent disease, and approximately 2000 patients have received this treatment, mostly for colorectal cancer, head and neck cancer and gynecologic cancer.
Low Energy IORT (50 kV)
Intrabeam,[14] (Carl Zeiss AG, Germany) received FDA and CE approval in 1999 and is a miniature and mobile X-ray source which emits low energy X-ray radiation (max. 50 kV) in isotropic distribution. Due to the higher ionization density caused by soft X-ray radiation in the tissue, the relative biological effectiveness (RBE) of low-energy X-rays on tumor cells is higher when compared to high-energy X-rays or gamma rays which are delivered by linear accelerators.[15] The radiation which is produced by low energy mobile radiation systems has a limited range. For this reason, conventional walls are regarded sufficient to stop the radiation scatter produced in the operating room and no extra measures for radiation protection are necessary. This makes IORT accessible for more hospitals. Targeted intra-operative radiotherapy is a low energy IORT technique.
See also
- Intraoperative Electron Radiation Therapy (IOERT)
- Brachytherapy
- External beam radiotherapy (EBRT) EBRT
References
- ↑ Mirnezami, Reza; Chang, George J.; Das, Prajnan; Chandrakumaran, Kandiah; Tekkis, Paris; Darzi, Ara; Mirnezami, Alexander H. (March 2013). "Intraoperative radiotherapy in colorectal cancer: Systematic review and meta-analysis of techniques, long-term outcomes, and complications". Surgical Oncology. 22 (1): 22–35. doi:10.1016/j.suronc.2012.11.001. PMID 23270946.
- ↑ Kraus-Tiefenbacher, Uta; Bauer, Lelia; Scheda, Antonella; Schoeber, Carola; Schaefer, Joerg; Steil, Volker; Wenz, Frederik (14 September 2007). "Intraoperative radiotherapy (IORT) is an option for patients with localized breast recurrences after previous external-beam radiotherapy". BMC Cancer. 7 (1). doi:10.1186/1471-2407-7-178. PMID 17854511.
- ↑ Hannoun-Levi, Jean-Michel; Ihrai, Tarik; Courdi, Adel (November 2013). "Local treatment options for ipsilateral breast tumour recurrence". Cancer Treatment Reviews. 39 (7): 737–741. doi:10.1016/j.ctrv.2013.02.003. PMID 23465859.
- ↑ Smyth, Chris (2014). "Single-dose radiotherapy eases breast cancer stress". The Times. Retrieved 26 July 2016.
- ↑ "Diagnosis and Treatment of Patients with Primary and Metastatic Breast Cancer" (PDF). AGO. Retrieved 26 July 2016.
- ↑ Belletti, B.; Vaidya, J. S.; D'Andrea, S.; Entschladen, F.; Roncadin, M.; Lovat, F.; Berton, S.; Perin, T.; Candiani, E.; Reccanello, S.; Veronesi, A.; Canzonieri, V.; Trovo, M. G.; Zaenker, K. S.; Colombatti, A.; Baldassarre, G.; Massarut, S. (3 March 2008). "Targeted Intraoperative Radiotherapy Impairs the Stimulation of Breast Cancer Cell Proliferation and Invasion Caused by Surgical Wounding". Clinical Cancer Research. 14 (5): 1325–1332. doi:10.1158/1078-0432.CCR-07-4453. PMID 18316551.
- ↑ Comas C., Prio A. Irradiation roentgen intra-abdominale ,après intervention chirurgicale dans un cas de cancer de l’uterus, Congres International d’Electrologie .Imprenta Francesca Badia,Barcelona,pp 5-14, 1907
- ↑ Comas, C; Prio, A (1906). Comptes rendus des séances du 3e Congrès international d'électrologie et de radiologie médicales (Milan: 5-9 septembre 1906) (in French). Lille: C. Robbo. p. 321.
- ↑ Abe, Mitsuyuki; Takahashi, Masaji (July 1981). "Intraoperative radiotherapy: The japanese experience". International Journal of Radiation Oncology•Biology•Physics. 7 (7): 863–868. doi:10.1016/0360-3016(81)90001-8. PMID 7198109.
- ↑ Goldson A., Past, present and prospects of intraoperative radiotherapy (IOR). Semin Oncol 1981.
- ↑ Vaeth, J. M. (1996). "Intraoperative Radiation Therapy in the Treatment of Cancer". Frontiers of Radiation Therapy and Oncology. 31. doi:10.1159/000061147.
- ↑ Abe M. History of Intraoperative radiation therapy. In:Debelbower RR, Abe M (eds) Intraoperative radiation therapy. CRC, Boca Raton; :1-10, 1989.
- ↑ Furhang, Eli E.; Sillanpaa, Jussi K.; Hu, Kenneth S.; Harrison, Louis B. (2011). "HDR-IORT: Physics and Techniques". In Gunderson, Leonard L.; Willett, Christopher G.; Calvo, Felipe A.; Harrison, Louis B. Intraoperative irradiation : techniques and results (2nd ed. ed.). New York: Humana Press. ISBN 978-1-61779-014-0.
- ↑ "INTRABEAM - breast cancer - Oncology - Medical Technology | ZEISS United States". Zeiss.com. Retrieved 2016-05-18.
- ↑ Hill, M. A. (15 December 2004). "The variation in biological effectiveness of X-rays and gamma rays with energy". Radiation Protection Dosimetry. 112 (4): 471–481. doi:10.1093/rpd/nch091.
External links
- Intraoperative radiation therapy at Cancer Treatment Centers of America
- Internal radiation therapy, cancer.org
- Interoperative radiation therapy at Mayo Clinic
- a mobile electron linear accelerator for IORT
- Report of AAPM Radiation Therapy IORT Committee Task Group No. 72