Orthovoltage X-rays

Orthovoltage X-rays
Intervention
ICD-9 92.22
OPS-301 code 8-521
A 200 kV orthovoltage x-ray tube used for radiation therapy, 1938. Orthovoltage x-ray machines are similar to diagnostic (radiography) x-ray machines, except that higher voltages are used and the x-ray tube is longer, to prevent the high voltages from arcing across the tube.

Orthovoltage x-rays are produced by x-ray tubes operating at voltages in the 200500 kV range, and therefore the x-rays have an energy in the 200500 keV range (see external beam radiotherapy for an explanation of the maximum and mean energies as a function of voltage). They have more energy than the x-rays used for diagnostic radiography, and are used in medicine in external beam radiotherapy to treat cancer and tumors. They penetrate tissue to a useful depth of about 46 cm.[1] This makes them useful for treating skin, superficial tissues, and ribs, but not for deeper structures such as lungs or pelvic organs.

Orthovoltage X-rays are sometimes termed "deep" x-rays (DXR). The energy and penetrating ability of the x-rays produced by an x-ray tube increases with the voltage on the tube. External beam radiotherapy (teletherapy) began around the turn of the 20th century with ordinary diagnostic x-ray tubes, which used voltages below 150 kV. Physicians found that these were adequate for treating superficial tumors, but not tumors inside the body. Since these low energy x-rays were mostly absorbed in the first few centimeters of tissue, to deliver a large enough radiation dose to buried tumors would cause severe skin burns.

Therefore beginning in the 1920s "orthovoltage" 200500 kV x-ray machines were built. These more penetrating x-rays were termed "deep" or "skin sparing" rays. These were found to be able to reach shallow tumors, but to treat tumors deep in the body more voltage was needed. By the 1930s and 1940s megavoltage X-rays produced by huge machines with 3-5 million volts on the tube, began to be employed. With the introduction of linear accelerators in the 1970s, which could produce 4-30 MV beams, orthovoltage x-rays are now considered quite shallow.

By convention, the voltage is used to characterize X- and gamma-ray beams (in volts), while electron beams are characterized by their energies (in electronvolts).

See also

References

  1. Advances in kilovoltage x-ray beam dosimetry http://iopscience.iop.org/0031-9155/59/6/R183/article
This article is issued from Wikipedia - version of the 7/14/2016. The text is available under the Creative Commons Attribution/Share Alike but additional terms may apply for the media files.